Chapter 12 Configuration and Test 26
12.2.6 Heating Parameters
Figure 12: Heating Parameters
Setpoint temperature: Temperature to which the
Ventus
is heated
default: 15°C
Heater mode: off: heating is off
auto: automated control to setpoint temperature
testmode:
Ventus
is heated to test setpoint temperature
ctrl high: heating off if control line on high,
otherwise: automatic control
ctrl low: heating off if control line on low
otherwise: automatic control
default: auto
Capacity: full: full heating power
alternating: upper plate and lower plate alternatingly
default: full
Disable if overvoltage:
the heating is disabled if overvoltage (28V) is detected
default: yes
Disable if undervoltage:
the heating is disabled if undervoltage (20V) is detected
default: no
12.2.7 Wind Parameters
Figure 13: Wind Parameters
Min. speed: starting speed of the wind sensor in m/s
default: 0,1m/s
Counts of measurements:
Number of measured values used for determining the statistical values
of average, minum value and maximum value
default: 60
Direction Offset: Correction of the wind direction
It is possible to have an offset added to the measured wind direction,
in order to convert e.g. 0° (North) into 180° (South)
default: 0°
Measurement rate: Update interval, adjustable in 250ms, 1 … 10 s
default: 10s
12.2.8 Pressure Parameters
Pressure offset: Absolute offset (for on-site adjustment) on the measured value