4. STROBOSCOPE MEASURING
PROCEDURES
4-1 Preparation and operating consideration
1)Install the handle " Handle " ( 3-13, Fig. 1 ) to
main instrument body, refer to
" Handle fix knot " ( 3-10, Fig. 2 ),
" Handle fix screw " ( 3-11, Fig. 2 )
" Handle fix knob " ( 3-12, Fig. 2 )
2)Connect the Power cable into the Power input socket
" 3-9, Fig. 1 ".
Caution :
* The power plug should be
connected to the correct AC
power supply.
3)A " Voltage rating label " is affixed under the
bottom case to show the voltage rating of
power supply.
When use the stroboscope, make sure to identify
the power supply voltage exactly.
Plug power cable's plug into a properly grounded
AC outlet.
4)For prolong life and safety, please adhere to the
following operation duty cycle:
< 2000 RPM - 2 hours
2001 to 3600 RPM - one hour
3601 to 8000 RPM - 30 minutes
> 8000 RPM - 10 minutes.
* 10 min. cooling off period between cycles.
6