SERVICE AND SPECIAL FUNCTIONS
MCB Types:
LiteOn:
C6/C7
Delta: C3
predefined limits,“0146 speedup skip”or“0146 steady
skip”, reduce the console command speed by the related
“0146 speed drop” values.
10.1MPH - Max, drop speed by 1
mph 6.1MPH - 10, drop speed by 1
mph
0.5 - 6 MPH, drop speed by 0.5 mph
When the counts of the “0144 OC warning” reach the
predefined limits,“0144 speedup skip”or“0144 steady
skip”, reduce the console command speed by the related
“0144 speed drop” values.
10.1MPH - Max, drop speed by 2
mph 6.1MPH - 10, drop speed by 1
mph
0.5 - 6 MPH, drop speed by 0.5 mph.
Counter limit for the “0146 OC warning” during the speed-
up process. By default, the console will not reduce the speed
command until the counter for the “0146 OC warning”
reaches 5.
Counter limit for the “0144 OC warning” during the speed-
up process. By default, the console will not reduce the speed
command until the counter for the “0144 OC warning”
reaches 2.
Counter limit for the “0146 OC warning” after the TM
reaches the steady-state. By default, the console will not
reduce the speed command until the counter for the “0146
OC warning”
reaches 2.
Counter limit for the “0144 OC warning” after the TM
reaches the steady-state. By default, the console will not
reduce the speed command until the counter for the “0144
OC warning”
reaches 2.