9 Maintenance
9.1 Regular Maintenance
9.1.1 Maintenance Plan
To ensure safe use and lengthen the service life of the video laryngoscope,
please conduct regular maintenance and check. Table 9-1 lists the maintenance
plan.
Table 9-1 Maintenance plan
Maintenance
Item
Frequency
Operator
Maintenance
Method
Appearance
check
Before each use
User
See section 9.1.2.
Power cable
check
Before each use
User
See section 9.1.3.
Electrical
safety test
Every two years
Distributor/Ho
spital engineer
See section 9.1.4.
Built-in battery
check
Every two years
Distributor/Ho
spital engineer
See section 9.2.3.
9.1.2 Appearance Check
Appearance check: Check that no crack or damage exists.
Button operation: Check that the buttons can be smoothly pressed and function
properly.
Check that all the sealing parts and the installation of the video laryngoscope
are normal and no crack exists on any materials.