Guidance and manufacturer’s declaration - electromagnetic immunity
The VS-10S/VS-10M/VS-10H video laryngoscope is intended for use in the
electromagnetic environment specified below. The customer or the user of the
VS-10S/VS-10M/VS-10H video laryngoscope should assure that it is used in such an
environment.
Immunity
Test
IEC 60601
Test level
Compliance level
Electromagnetic
environment -guidance
Conducted
RF
IEC61000-4-
6
3 Vrms
150 kHz to
80MHz;
6 Vrms in ISM
bands
a
Between
0.15MHz and 80
MHz;
80% AM at 1
kHz
3 Vrms
150 kHz to
80MHz;
6 Vrms in ISM
bands
Between 0.15MHz
and 80 MHz;
80% AM at 1 kHz
Portable and mobile RF
communications
equipment should be
used no closer to any
part of
theVS-10S/VS-10M/VS
-10H video
laryngoscope, including
cables, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
Pd 2.1
Radiated RF
IEC61000-4-
3
3 V/m
80 MHz 2.7
GHz;
80% AM at 1
kHz
27V/m:380-390
MHz;
28V/m:430-470
3 V/m
80 MHz 2.7
GHz;
80% AM at 1 kHz
27V/m:380-390M
Hz;
28V/m:430-470M
Hz;