or relocating the VS-10S/VS-10M/VS-10H video laryngoscope.
d Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
[3] V/m
Recommended separation distances between portable and mobile RF communications
equipment and the VS-10S/VS-10M/VS-10H video laryngoscope
The VS-10S/VS-10M/VS-10H video laryngoscope is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the VS-10S/VS-10M/VS-10H video laryngoscope can help
prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
VS-10S/VS-10M/VS-10H video laryngoscope as recommended below, according to
the maximum output power of the communications equipment.