quick set up guide
1 Remove existing mattress and replace with
ProCair Safety Underlay (printed top cover
upwards; air hoses towards bed base).
2 Place overlay on top of
Underlay (printed top cover
upwards; air hoses towards
bed base).
Slide elastic straps over mattress ends to secure.
Ensure straps fit firmly and are not twisted or pulled
towards corners.
If not using the ProCair Safety Underlay, place overlay
on top of your existing mattress and secure with elastic.
3 Hang the control unit using
the rear hanging hooks or
place on a stable surface (ie
floor beneath the bed).
4 Attach air hoses to the control
unit. Straighten any twists to
ensure uninterrupted air flow
between control unit and
overlay.
5 Ensure the CPR tag sealing
connector is fully inserted
into air hoses (tag is located
on side, near head end)
6 Feed the power cord through
retention loops on side of
overlay to keep cord off the
floor.
7 Insert the power cord into
control unit base, then plug
into power socket and switch
on mains power.
8 Press the Power button
for two (2) seconds to
switch on.
The Pressure Setting
indicators (Soft to Hard)
flash to indicate the system
is operational and inflating.
Allow up to 40 minutes
for complete inflation.
When inflation is complete,
the fourth indicator, together
with Alternating Mode
indicator will illuminate.
9 Once the overlay is inflated, bedding can be
replaced. Secure sheets loosely enough to ensure
they do not interfere with cell alternation.
10 Lie patient face up
and test pressure settings
by performing a
‘bottoming out’ test:
a unzip one side of the top cover
b slide your hand underneath the patient, feeling
for a deflated cell under the patient’s lower
spine – your hand should easily slide between
patient and base
c for optimal pressure: patient should be
immersed 30-50% into fully inflated cells
d if patient is not appropriately immersed, decrease
pressure setting and repeat test every 10-12
minutes, until optimal pressure is reached
DIGITAL
1
2
3
less pressure, more support
freecall 1300 882 194
|
freefax 1300 882 197
|
www.medistore.com.au
May15
Question and Answer IconNeed help?

Do you have a question about the Medistore ProCair Digital 5000 and is the answer not in the manual?

Medistore ProCair Digital 5000 Specifications

General IconGeneral
BrandMedistore
ModelProCair Digital 5000
CategoryMedical Equipment
LanguageEnglish

Summary

ProCair 5000 Quick Setup Guide

Initial Setup and Connection

Connect underlay, overlay, air hoses, and power cord to the control unit, then plug into the mains.

System Power-On and Inflation

Press the power button to turn on the system; allow up to 40 minutes for complete inflation.

Patient Comfort and Testing

Perform a 'bottoming out' test to ensure optimal pressure settings for patient comfort and immersion.

Control Unit Operation and Features

Control Unit Button Functions

Details the functions of the power, mute, comfort control, mode selection, max inflate, and lockout buttons.

Alarm Indicators and Fault Diagnosis

Explains visual and audible alarm signals, identifying potential faults like power issues or pressure errors.

Transport Mode Procedure

Instructions for preparing the unit for transport, including setting static mode and securing air hoses.