LOT: XXXXXX ID: XXXX
P/N: X-X-XXXHAN
2-8-611 HOOKUP CONNECTION GUIDE
Using MEDIVATORS
®
ADVANTAGE
®
Automated Endoscope Reprocessors
Port ID Port Name Connector Assembly Description Order Number
1
LEAK TEST Leak Test Connector Assembly H28611-1
2
JET/AUX Air/Water Connector Assembly H28611-26
3
BIOPSY Biopsy Connector Assembly H28611-3
4
AIR Air Probe Connector Assembly H28611-4
5
ELEVATOR Elevator Wire/Auxiliary Water Connector Assembly H28611-5
6
WATER Air/Water Connector Assembly H28611-26
7
SUCTION Suction Connector Assembly H28611-7
A
NA Air/Water & Suction Port Separator 1-5-112
If a Connector Assembly needs replacement, locate the port number/name and use
the corresponding order number found in the Connector Assembly Identication Chart.
Connector Assemblies are specic to individual hookups. Check each Hookup Connection
Guide for relevant information.
ADVANTAGE HOOKUP and CHANNEL SEPARATOR (Figure 1)
Each ADVANTAGE Hookup is comprised of multiple ports and connector assemblies.
Each port is labeled with a Port ID number (circled). This same number also identies the
corresponding Connector Assembly. Each Connector Assembly is tted with a specic
endoscope connector. (see Fig. 2).
EXAMPLE CONNECTOR ASSEMBLY (Figure 2)
READ ALL INSTRUCTIONS PRIOR TO CONNECTION
Locate the endoscope port* and follow the instructions for attaching the
corresponding Endoscope Connector. All connectors must be attached.
CONNECTOR ASSEMBLY IDENTIFICATION CHART
Ferrule
Endoscope
Connector
example
Tubing
Connector
Coupling
Label
A

Leak Test Connector
Align the slot on connector with the pin on the
scope. Push on and turn clockwise to engage.
To remove leak test connector, turn connector
counter-clockwise until slot is aligned with pin and
pull off. Do not pull until slot is aligned with pin.
Suction Connector
Align silicone tubing with
suction barb and push tubing
onto barb until fully engaged.
Elevator Wire Channel Connector
(endoscope model dependent)
Align the connector with the endoscope tting
and screw clockwise until fully engaged. (Check
alternate location ve)
Air/Water and Suction
Port Separator
Align the separator with
piston ports and push to
compress springs before
sliding the frame forward to
fully engage.
A
Air Probe Connector
Align with Air Probe, push and
slide until fully engaged.
Biopsy Connector
Align the biopsy
connector disc over the
biopsy inlet and push
down to compress spring
and slide the frame
forward to fully engage.
Air/Water Connector
Align the connector
openings with the air/water
ports of the endoscope as
shown. Ensure side latch
is in. Push connector onto
ports until latch moves
out and clicks. To remove,
push latch to release.

Auxiliary Water Connector
(endoscope model dependent)
Align the connector with the endoscope
tting and screw clockwise until fully
engaged. (Check alternate location ve)
* Representative endoscope for illustrative purposes only. For exact
port locations, refer to endoscope manufacturer’s instruction guide.
Ensure proper alignment of all connectors with the
endoscope mating parts prior to attachment.
If you have any questions regarding this hookup, in the U.S., please contact MEDIVATORS Technical Support at 1-800-444-4729. Outside the U.S., please contact your MEDIVATORS Distributor.
DESIGNATED HOOKUP FOR THE FOLLOWING OLYMPUS
®
ENDOSCOPES*:
CF-1T140I CF-H185I CF-HQ290L CF-Q160L GIF-1T140L GIF-HQ190 JF-1T40 PCF-Q180AL TJF-145 TJF-M20
CF-1T140L CF-H185L CF-Q145I CF-Q160S GIF-1T160 GIF-HQ290 JF-240 PJF-16 TJF-160F
CF-FH260AZI CF-H190I CF-Q145L CF-Q165I GIF-1T240 GIF-XK240 PCF-H180AI PJF-160 TJF-160R
CF-FH260AZL CF-H190L CF-Q160 CF-Q165L GIF-1TQ160 GIF-XTQ160 PCF-H180AL PJF-240 TJF-160VF
CF-H180AI CF-HQ190I CF-Q160AI CF-Q180AI GIF-H180J JF-140 PCF-H190I TJF-140 TJF-160VR
CF-H180AL CF-HQ190L CF-Q160AL CF-Q180AL GIF-H190 JF-140F PCF-H190L TJF-140F TJF-240
CF-H80DI CF-HQ290I CF-Q160I GIF-1T140 GIF-H190J JF-140R PCF-Q180AI TJF-140R TJF-260V
* This list of corresponding endoscope models is subject to change without notice. For a full list, please go to: www.medivators.com/hookuplookup