Veris
TM
8600
Vital Signs Monitor
Operation Manual
3010796 Revision 0
Date 10/04
PRELIMINARY

Table of Contents

Questions and Answers

  • B
    Benjamin WoodsAug 17, 2025
    Why is there ECG noise on my Medrad Measuring Instruments?
    • V
      Vanessa HernandezAug 17, 2025
      If the ECG on your Medrad Measuring Instruments is showing noise or is intermittently malfunctioning, first, try re-prepping the patient's skin. If that doesn't work, reposition or replace the electrodes. Also, check the ECG settings to ensure the line frequency/filter is correct. If the problem persists, the fiber optic cable or the ECG module may be defective and need replacement.
  • B
    Bradley KennedyAug 20, 2025
    What to do if the remote display of my Medrad Veris 8600 Measuring Instruments won’t power up?
    • T
      Thomas LloydAug 20, 2025
      If the remote display of your Medrad Measuring Instruments won’t power up and the On/Off LED is dark, make sure the AC power cord is securely connected to the monitor and plugged into a live outlet. Also, check the fuses located above the AC connection and replace them if they are blown.
  • J
    Joshua ReedAug 22, 2025
    How to fix a Medrad Measuring Instruments main monitor that won’t power up?
    • S
      Shawn SanchezAug 23, 2025
      If the main monitor of your Medrad Measuring Instruments won’t power up and the On/Off LED is dark, the batteries may be discharged. Connect the monitor to the power supply to recharge the batteries.
  • S
    Samuel GomezAug 26, 2025
    What to do if the display is blank on my Medrad Veris 8600 Measuring Instruments?
    • B
      brennanjamesAug 26, 2025
      If the display on your Medrad Measuring Instruments is blank or not readable, it could be due to discharged batteries. Connect the monitor to the power supply to recharge them. Also, check that the AC power cord is securely connected to the remote display.
  • L
    Lisa KelleyAug 28, 2025
    How to troubleshoot NIBP air leak in Medrad Veris 8600 Measuring Instruments?
    • D
      daniel49Aug 28, 2025
      If you are experiencing an NIBP air leak with your Medrad Measuring Instruments, the cuff or hose may be defective. Try replacing the cuff first, and if the issue continues, replace the hose. If neither of these steps resolves the leak, further assistance is needed.
  • J
    James LittleAug 31, 2025
    What to do if NIBP is not functioning on Medrad Veris 8600 Measuring Instruments?
    • L
      lucasrichardAug 31, 2025
      If the NIBP is not functioning on your Medrad Measuring Instruments, first ensure that the NIBP cable and extension cable are securely connected.
  • A
    angela23Sep 2, 2025
    How to fix SpO not functioning on Medrad Veris 8600?
    • K
      Kristi MooreSep 3, 2025
      If the SpO is not functioning on your Medrad Measuring Instruments, ensure that the SpO cable and extension cable are securely connected.
  • J
    jerrybeltranSep 5, 2025
    Why is there no sound from the speaker on my Medrad Measuring Instruments?
    • B
      brian10Sep 5, 2025
      If there is no sound coming from the speaker of your Medrad Measuring Instruments, the volume may be turned down. Adjust the volume to a higher level in the Configuration window.
  • L
    Laura WoodSep 8, 2025
    How to fix COM 1 serial interface on Medrad Veris 8600?
    • E
      evanjosephSep 8, 2025
      If the COM 1 serial interface is not working on your Medrad Measuring Instruments, it may be set to the internal printer. Reset the monitor to the external printer.
  • S
    Sarah AlvaradoSep 12, 2025
    What causes a trap error on Medrad Veris 8600?
    • S
      Scott CookSep 12, 2025
      If you encounter a trap error or CO2 malfunction on your Medrad Measuring Instruments, the trap may be plugged or leaking. Replace the trap. Also, check if the O Cell is loose and tighten it.

Summary

2 - Monitor Setup

Battery Power

Discusses the monitor's battery operation and charging.

System Start and Auto-calibration

Describes the process of powering up the monitor and initial auto-calibration.

ALARMS Softkey

Accesses parameter alarm settings windows.

PARAMS Softkey (Physiological Parameters)

Accesses physiological parameter settings for modules.

DISPLAY Softkey

Adjusts display settings, including waveform configurations.

CONFIG Softkey (System Configuration)

Accesses system configuration settings.

PRINT Softkey

Accesses the printer settings window.

Default Settings

Lists factory default alarm settings for various parameters.

3 - Alarms and Messages

Alarm Description

Explains how the monitor alerts operators to status changes and parameter alarms.

Alarm Silence

Details the function of the SILENCE key for temporary or permanent alarm muting.

5 - ECG

ECG Monitoring (Electrocardiogram)

Provides instructions for preparing a patient for ECG monitoring.

Gating Interface

Describes how to connect for gating using ECG or SpO2 signals.

6 - NIBP

NIBP Monitoring

Provides instructions for preparing a patient for NIBP monitoring.

Procedure

Outlines the steps for performing an NIBP measurement.

7 - SpO2

SpO2 Monitoring Procedures (Pulse Oximetry)

Provides instructions for preparing a patient for SpO2 monitoring.

SpO2 Peripheral Gating

Describes using SpO2 signal as an alternative to ECG gating.

8 - IBP

Theory of Operation

Explains IBP monitoring using intra-arterial catheters and transducers.

IBP Monitoring

Describes real-time invasive blood pressure monitoring.

Setup and User Calibration

Describes the process of periodically calibrating the IBP monitor.

10 - Anesthetic Agents

Theory of Operations

Explains integrated CO2 and agent gas detection.

Gas Monitoring Safety

Outlines precautions and contraindications for gas monitoring.

Appendix B: Unit and Configuration Defaults

Configuration Default Settings

Explains how to install different configuration defaults (Base, Cardiac, Anesthesia).

ALARMS Menu Settings

Details default alarm settings for different configurations.

Appendix C: Specifications

Alarms

Lists alarm characteristics, levels, settings, and modes.

Appendix E: Troubleshooting

General Troubleshooting

Provides general checks for monitor and accessory placement and connections.

Troubleshooting Table

Lists common symptoms, problems, and solutions for monitor malfunctions.

Medrad Veris 8600 Specifications

General IconGeneral
DisplayColor LCD
Power Supply100-240V AC, 50/60 Hz
Wavelength RangeNot applicable
Battery LifeUp to 8 hours
ConnectivityWireless
CompatibilityMRI systems
Power RangeNot Applicable