1
Table of Contents
PAGE
Purpose and Scope of This Document ........................................................1
Returning a Temporary Pacemaker to Medtronic .............................................2
5392 Control Description ....................................................................2
Test Parameter Definitions ..................................................................3
Test Conditions . ............................................................................4
Test Equipment .............................................................................4
Test Procedures .............................................................................5
1) Rate ........................................................................................5
2) Rapid Atrial Pacing Test (RAP) ...............................................................5
3) Output. ..................................................................................... 6
4) Pulse Width .................................................................................7
5) A/V Interval .................................................................................8
6) Sensitivity ..................................................................................8
7) Emergency .................................................................................9
8) Inactive Current Drain ....................................................................... 9
9) Active Current Drain .......................................................................10
10) Battery Removal – Pacing Continuation Test ...............................................10
Error Buttons and Messages ................................................................11
5392 Service Life Statement ................................................................. 11
Returned Product Information Report .......................................................
12
Purpose and Scope of This Document
This manual is intended for use by a qualified engineer or technician for the purpose of checking the
operation of a Medtronic Model 5392 Dual Chamber Temporary Pacemaker. The instructions in this
manual cover testing the electrical parameters of the Model 5392 temporary pacemaker. If checkout
of a temporary pacemaker indicates improper operation, the unit should be returned to Medtronic for
evaluation and repair.
Medtronic does not recommend field repair of this device. Refer to terms of the limited warranty covered
in the Model 5392 Dual Chamber Temporary Pacemaker Technical Manual supplied with the device.
Note: There is no method to adjust an out-of-tolerance parameter at an off-site location. The device must
be returned for recalibration.
Refer to the Model 5392 Dual Chamber Temporary Pacemaker Technical Manual for a complete
description, operating instructions, cleaning and maintenance instructions, and specifications for this
device.
Model 5392
Dual Chamber Temporary Pacemaker Checkout Manual