Step 1: Prepare the capsule delivery device
1) Check the expiration date on the capsule.
2) Without bending or kinking the delivery device, carefully remove it from the external shipping box, and
then from the inner pouch (Fig. 1).
3) Remove the capsule’s plastic cover, and magnetic clip. The capsule will automatically turn on when the
magnetic clip is removed.
4) Set the magnetic clip aside (Fig. 2).
5) After opening the capsule package, make sure the capsule trocar needle is not
advanced (Fig.3).
Step 2: Prepare the vacuum pump
550 mmHg. Make a note of the gauge reading (Fig. 4).
50 mmHg to 500 mmHg or lower.
Step 3: Start the recording
2) From the main menu, select Start Study and press Enter.
3) If you see a “Last study not uploaded!” message, it means the recorder has detected existing data from
a previous study that has not been uploaded.
4) To stop and upload the existing data, press Cancel and connect to the PC. Follow the instructions on
the recorder. When the data is uploaded, start the new study again by selecting Start Study from the
main menu.
5) To continue (overwrite the existing data without uploading it), press Next and then Yes
6) The recorder will start searching for the capsule's signal, and the “Clearing data...” message will appear,
followed by “Waiting for pH capsule 1.”
7) When you see the message,” Does this pH capsule ID number match the delivery device number?”
Press Yes or No
™
™
begin again.
8) If the recorder still does not recognize the capsule or displays a capsule mismatch error, repeat
the procedure.
9) If this is a two-capsule study, repeat the process.
10) When capsule recorder pairing is complete, the recorder will automatically start recording
and the screen will display the symptom buttons.
is blinking blue.
Open the
plastic cover
Remove
magnetic
clip
Suction
chamber
Trocar
needle
Correct Incorrect
A. Connect
vacuum
hose
B. Place
finger
over
suction
chamber
C. Remove
finger
vacuum
hose
vacuum
port
mmHg
700
600
500
400
300
200
100
0
400
mmHg
700
600
500 300
200
100
0
Figure 1
Figure 2
Figure 3
Figure 4
™
Quick reference guide