Alarms and alerts in SmartGuard Auto Mode
In addition to the pump and sensor alarms that were introduced in GETTING STARTED WITH MINIMED™
770G INSULIN PUMP, you may receive the following alerts and alarms only when the pump is in
SmartGuard Auto Mode.
Title and text Cause Next Steps
Auto Mode started
The following
SmartGuard settings are
now turned o:
• Suspend before low
• Suspend on low
Your pump has started Auto Mode. The
SmartGuard Suspend before low and Suspend
on low settings are now turned o.
• Select OK to clear
the alert.
• Alert is information
only. No action is
required at this time.
Auto Mode Exit
Basal Name started.
Would you like to
review the Auto Mode
Readiness screen?
Your pump has exited Auto Mode. Follow instructions
on the pump screen.
Check the Auto Mode
Readiness screen for
information to re-enter
Auto Mode.
Auto Mode max
delivery
Auto Mode has been
at maximum delivery
for 4hours. Enter BG to
continue in Auto Mode.
Alerts you when your pump has been delivering
insulin at your maximum Auto Mode basal
delivery rate for 4hours. Your personal maximum
Auto Mode basal delivery rate is automatically
determined.
• Select OK to clear
the alert.
• Enter a BG to
continue in Auto
Mode.
Quick Reference Guide
Alerts & Alarms in Auto Mode
MINIMED
TM
770G