Installation and User
Guide
NovAseptic
®
GMP ATEX Mixer
NovAseptic and the Vibrant M are trademarks of Merck KGaA, Darmstadt, Germany or its aliates. All other trademarks are the
property of their respective owners. Detailed information on trademarks is available via publicly accessible resources. © 2008-
2018 Merck KGaA, Darmstadt, Germany and/or its aliates. All Rights Reserved.
00105442PU Rev 5 02/2018

Questions and Answers

  • W
    William WagnerAug 12, 2025
    What to do if the Merck Mixer motor/mixing head does not rotate?
    • K
      kimberlycamachoAug 13, 2025
      If the motor or mixing head of your Merck Mixer isn't rotating, first check the power supply to ensure it's properly connected. If the power is fine, the overload protection might have tripped; in this case, verify the amp value setting and the ambient temperature. Lastly, check for and remove any heavy buildup of particles stuck to the magnets, then clean the unit.
  • R
    Rachel WeaverAug 16, 2025
    Why is the outer drive head on my Merck EMD Millipore NovAseptic scratching the tank plate?
    • K
      Katherine ThompsonAug 16, 2025
      If the outer drive head of your Merck Mixer is scratching the tank plate, ensure the drive unit is correctly installed. If it is, check for and remove any particles stuck to the magnets. Also, review the tank plate welding instructions and verify tolerances with appropriate gauges to correct any deformation. If the shaft or hole-shaft in the gearbox is bent, or the outer drive head is loose, contact Technical Service.
  • E
    Emily WoodAug 20, 2025
    What causes magnetic disconnection of the mixing head in a Merck Mixer?
    • C
      clarkwesleyAug 20, 2025
      If the mixing head of your Merck Mixer is decoupling (magnetic disconnection), consider these factors: * Review the guidelines for starting the mixer to ensure the correct order of substance addition and avoid dry running. * Inspect male bearing and tank plate for damage. If it is undamaged, reinstall the bearing and check rotation direction of the mixing head. * Increase the ramp time. * Remove any particles stuck to the magnets.
  • D
    dominique64Aug 25, 2025
    What causes the mixing head of my Merck EMD Millipore NovAseptic to scratch against the tank plate?
    • M
      Matthew MullinsAug 25, 2025
      If the mixing head of your Merck Mixer is scratching against the tank plate, it could be due to several reasons. The male bearing might be worn out and need replacement. Alternatively, the female bearing holder could be damaged, requiring repair of the mixing head. If the male bearing is loose, remove it, inspect for damage, and if undamaged, reinstall, checking the mixing head's rotation direction. Also, remove any particles stuck to the magnets. Finally, review the tank plate welding instructions and verify tolerances to correct any deformation.
  • C
    Crystal CooperAug 27, 2025
    Why does my Merck Mixer have a high amp value?
    • C
      chavezlindseyAug 27, 2025
      If your Merck Mixer is showing a high amp value, the ambient temperature might be too high, so try chilling the environment. Also, the motor could be mechanically overloaded, so check all transmissions.
  • T
    Tracy SmithAug 31, 2025
    Why is the mixing head on my Merck EMD Millipore NovAseptic Mixer running eccentrically?
    • R
      Roberta MillerSep 1, 2025
      If your Merck Mixer's mixing head is running eccentrically, the male bearing might not be properly placed. Remove the male bearing, inspect it and the tank plate for damage. If there's no damage, reinstall the bearing, checking the rotation direction of the mixing head. Alternatively, the bearings could be worn out, requiring replacement of the male bearing or repair of the mixing head.
  • B
    bakerfrankSep 4, 2025
    What causes an abnormal noise in my Merck Mixer?
    • T
      Terry GilmoreSep 4, 2025
      If your Merck Mixer is making an abnormal or rumbling noise, it could be due to a deep vortex. Try reducing the speed and ensure the mixing head is fully covered with liquid at all times. Cavitation might also be the cause, so investigate the pressure, temperature, and speed. Another possible cause is dry running; in that case, review the guidelines under “Operation /Starting the Mixer.”
  • D
    Donna MorganSep 8, 2025
    Why is my Merck EMD Millipore NovAseptic Mixer losing running speed?
    • K
      Kristen MartinezSep 8, 2025
      If your Merck Mixer is losing running speed, start by checking the power supply to ensure it's correct. Also, verify the fuses to make sure no phases are missing (AC).

Summary

Installation Note

Operator and Equipment Safety

NovAseptic GMP ATEX Mixer Components

Installation and Operating Parameters

Installation

Male Bearing Installation

Step-by-step guide for installing the male bearing component.

Mounting the Male Bearing

Detailed steps and torque specifications for mounting the male bearing.

Mixing Head Installation

Drive Unit Installation

Precautions and prerequisites for installing the drive unit.

Check Direction of Rotation

Mounting the Drive Unit

Instructions for mounting the drive unit onto the tank plate for different models.

Electrical Installation

Operation

Operating Notes

Cleaning and Sterilizing Procedures

Installation ATEX Checklist

Install the Male Bearing

Checklist items to verify correct male bearing installation.

Install the Mixing Head

Checklist items to verify correct mixing head installation.

Before Installing the Drive Unit

Checklist items to verify prerequisites before drive unit installation.

Install the Drive Unit - GMP 50 ATEX, GMP 100 ATEX, GMP 500 ATEX

Checklist items for installing specific drive unit models.

Install the Drive Unit - GMP 1000 ATEX and GMP 2000 ATEX

Install the Drive Unit - GMP 5000 ATEX, GMP 10000 ATEX, GMP 20000 ATEX

Checklist items for installing specific drive unit models.

Before Installing the Electrical Components

Checklist items to verify prerequisites before electrical component installation.

Install the AC Motor

Checklist items for AC motor installation verification.

Install the VFD System

Commissioning

Valid Regulations

Checklist to ensure ATEX regulations are met during commissioning.

Installation Qualification Checklist

Maintenance Checklist

Troubleshooting

Merck EMD Millipore NovAseptic Specifications

General IconGeneral
BrandMerck
ModelEMD Millipore NovAseptic
CategoryMixer
LanguageEnglish