Preparation for use | EN
13
> The device must be connected to the sewage pipe via an odour trap.
> Thewastewatermustbeabletoowfreely,astheseparationefciencyisnotguaranteedintheeventofbackwater.
> Water jet operated saliva ejectors must be shut down due to high water consumption.
> Theushingofthespittoonmustbelimitedbymeansofatimerorbuttontomax.30secondsatamax.waterowof3l/mintoenablethecentrifugetostop.
> Formatchinghoseadaptersandttingssee9.1.1.Accessories,servicekits,collectioncontainersandspareparts.
5.2.3. Connection dimensions
ø 15
ø 26
ø 15
ø 26
5.2.4. Filter
Theremustbeacoarselterinthedrainofthespittoon.Thismusthaveameshsizeofmax.3mmandmustnotberemovedwhiletheamalgamseparatorisin
operation.
Ifnotpresentinthehosestorage,apre-lterwithameshsizeofmax.1mmmustbeinstalledonthesuctionsideoftheamalgamseparator.Thepre-ltermust
be easily accessible for the practice staff.
1 Pre-lter
2 Coarselter(spittoon)