SAFETY INFORMATION
Read the instruction manual before use.
It is only allowed to open the instrument to change
the batteries.
We cannot be held liable for damage caused by
injudicious, incorrect or improper use of this
instrument.
We are not liable for consequential damage that
arises from this.
The instrument should not be exposed to high tem-
peratures, strong vibrations or high humidity.
Always keep out of reach of children or animals.
Keep the instrument dry.
Make sure that no dirt IS between the probes.
Changing the electrodes:
Unscrew the electrodes from the housing and
replace them
MAINTENANCE
SPECIFICATIONS
Measuring principle
Length electrodes
Electrodes
Wood measuring range
Construction material range
Accuracy
Batteries
Housing
Ambient temperature
Relative humidity
Dimensions
Weight
Guarantee
Electrical resistance
8mm
interchangeable
1 - 45% (±1%)
0.1 - 2.4% (± 0.1%)
± 3.0% RH
3x CR2032
Impact-resistant plastic
0 - 20°C
0 - 85%RH
139 x 47 x 25mm
100g
1 year
CONTROL CHECK
“O------T------O” Press the electrodes onto the T contacts of
the protective cap. Display value for wood: 27%±2% Value
for building materials: 1.3%±0.1%
“ O…… B……O” Press the electrodes onto the B contacts of
the protective cap. Display value for wood: 44%±2% Value
for building materials: 2.1%±0.1%
If a dierent value appears, clean the electrodes. If no
change occurs, the instrument needs to be checked.