QA-ST MKII – Getting Started Guide
5
Appendix A Specifications QA-ST MK II US
SUPPLY VOLTAGE MEASUREMENT
Range: 90 - 125V
Accuracy: ± (2 % of reading + 2 digits)
Resolution: 1 V
CURRENT CAPACITY
15 Amps @ 100 Volts (15 Amps limited duration)
15 Amps @ 120 Volts (15 Amps limited duration)
GROUND WIRE RESISTANCE MEASUREMENTS
• Test current: AC Source – 200 mA
• Test current accuracy: (+5 % of reading)
• Range: 0 - 1999 mΩ
• Accuracy: ± (2 % of reading + 10 mΩ)
• Resolution: 1 mΩ
LEAKAGE CURRENT
• Range: 0 – 1999 µA RMS
• Accuracy: DC and 25 to 1 kHz
± (2 % of reading + 3 µA)
• Resolution: 1 µA
PATIENT LEAKAGE CURRENT
MEASUREMENTS (DC AND AC)
• Range: 0 – 1999 µA
• Accuracy: DC and 25 to 1 kHz,
± (2 % of reading + 3 µA)
• Resolution: 1 µA
• Test modes: AC and DC measurements.
LEAD ISO MEASUREMENTS
• Range: 0 – 6000 µA RMS
• Accuracy: DC and 25 to 1 kHz,
± (2 % of reading + 3 µA)
• Resolution: 1 µA
INPUT IMPEDANCE
1000 ohm, AAMI ES1-1993 load or IEC 60601-1 load
POWER CONSUMPTION
110 V (50/60 Hz) = 200 mA
115 V (50/60 Hz) = 200 mA
GENERAL INFORMATION
Temperature Requirements:
+ 15/59 to + 35/95 °C/F while operating
0 to + 50/122 °C/F for storage
Display:
Type: LCD
Graphical: 122 x 32 pixels
Display control: Keypad
Power: 100 V 50/60 Hz
Housing: Plastic case
Dimensions: D x W x H
191mm x 135mm x 45mm
Weight: 930 g
Standard Accessories:
User Manual
Test lead
10 Universal snap to banana adapters