Manual Cleaning/Processing
Refer to the PAL-650 Instructions for Use (IFU) for
cleaning parameters, as any deviation from these
parameters may result in inadequate cleaning of
PAL Handpiece and cable.
Clean the Handpiece thoroughly with warm
(> 60 °C / 140 °F) water, pH neutral enzymatic
detergent, and a soft brush. Scrub the Handpiece
with a brush, paying close attention to
instrument crevices.
Rinse thoroughly under running (< 50 °C / 122 °F)
water for a minimum of 2 minutes. If possible,
use distilled water for the final rinse.
For more information, please reference PAL System Instructions for Use (IFU) “IM-PAL-650” and “IM-5025”
LIT-PAL-CS Rev. A 15