All sterilization cycles require a MINIMUM
DRY CYCLE exposure time of 20 MINUTES.
This is a minimum requirement and should
be increased accordingly until ALL evidence
of moisture have been removed from the
device and packaging.
Never shorten the dry cycle, this may result
in service interruptions related to residual
moisture and corrosion that may not be
visible during examination.
Always allow the PAL Handpiece and Cable
to cool slowly at room temperature.
Attempting use prior to cooling may result in
lost functionality and/or damage the system.
IFU Steam Sterilization
For more information, please reference PAL Cleaning & Sterilization Chart (IM-PAL-CS)
LIT-PAL-CS Rev. A 20