7BP A200
EN
Instrument care
Clean the instrument only with a soft, dry cloth.
Cleaning the cuff
Carefully remove spots on the cuff with a damp cloth and soapsuds.
Accuracy test
We recommend this instrument is tested for accuracy every
2 years or after mechanical impact (e.g. being dropped). Please
contact Microlife-Service to arrange the test (see foreword).
Disposal
11. Guarantee
This instrument is covered by a 5 year guarantee from the date
of purchase. The guarantee is valid only on presentation of the
guarantee card completed by the dealer (see back) confirming date
of purchase or the receipt.
Please contact Microlife-Service (see foreword).
12. Technical Specifications
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
WARNING: Do not wash the cuff in a washing machine or
dishwasher!
Batteries and electronic instruments must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.
Batteries, cuff and wearing parts are not included.
Opening or altering the instrument invalidates the guarantee.
The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
Operating tempera-
ture:
10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage temperature: -20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight: 560 g (including batteries)
Dimensions: 152 x 92 x 42 mm
Measuring proce-
dure:
oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V dias-
tolic
Measurement range: 20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range: 0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within ± 3 mmHg
Pulse accuracy: ± 5 % of the readout value
Voltage source: 4 x 1.5 V Batteries; size AA
Mains adapter DC 6V, 600 mA
(optional)
Reference to
standards:
EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife BP A200 and is the answer not in the manual?

Microlife BP A200 Specifications

General IconGeneral
BrandMicrolife
ModelBP A200
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals