7IFR 100 EN
10.Cleaning and Disinfecting
Use an alcohol swab or cotton tissue moistened with
alcohol (70% Isopropyl) to clean the thermometer casing
and the measuring sensor. Ensure that no liquid enters the
interior of the device. Never use abrasive cleaning agents,
thinners or benzene for cleaning and never immerse the
device in water or other cleaning liquids. Take care not to
scratch the surface of the sensor lens and the display.
11.Battery Replacement
This device is supplied with one lithium battery, type
CR2032. Replace with a new CR2032 battery when the
flashing battery symbol appears on the display BK.
Remove the battery cover BL by sliding it in the direction
shown. Insert the new battery with the + at the top.
Batteries and electronic devices must be dis-
posed of in accordance with the locally applicable
regulations, not with domestic waste.
12.Guarantee
This device is covered by a 2 year guarantee from the date
of purchase. The guarantee is valid only on presentation of
the guarantee card completed by the dealer (see back)
confirming date of purchase or the receipt.
The guarantee covers the device. The battery and pack-
aging are not included.
Opening or altering the device invalidates the guarantee.
The guarantee does not cover damage caused by
improper handling, a discharged battery, accidents or
non-compliance with the operating instructions.
Please contact Microlife-service.
13.Technical Specifications
Type:
Dual Mode Thermometer
IFR 100
Measurement
range:
Ear Mode:
0-100.0 °C / 32-212.0 °F
Forehead Mode:
34-42.2 °C / 93.2-108.0 °F
Resolution:
0.1 °C / °F
Measurement
accuracy:
Laboratory:
±
0.2 °C, 32.0
42.2 °C
±
0.4 °F, 89.6
108.0 °F
Display:
L
iquid
C
rystal
D
isplay, 4 digits plus special icons
Acoustic:
The unit is turned ON and ready for the
measurement: 1 short beep.
Complete the measurement: 1 long beep
(1 sec.) if the reading is less than 37.5 °C /
99.5 °F, 10 short «beep» sounds, if the reading
is equal to or greater than 37.5 °C / 99.5 °F.
System error or malfunction: 3 short «bi»
sounds.
Memory:
Auto-Display the last measured temperature
12 readings recall in the Memory Mode
Backlight:
The display light will be GREEN for 4 seconds,
when the unit is turned ON.
The display light will be GREEN for 5 seconds,
when a measurement is completed with a
reading less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds,
when a measurement is completed with a
reading equal to or higher than 37.5 °C /
99.5 °F.
Operating
conditions:
Ear Mode:
10-40.0 °C / 50-104.0 °F
Forehead Mode:
16-40.0 °C / 60.8-104.0 °F
Storage
conditions:
-25- +5C / -13- +13F
15-95
% relative maximum humidity

Other manuals for Microlife IFR 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microlife IFR 100 and is the answer not in the manual?

Microlife IFR 100 Specifications

General IconGeneral
BrandMicrolife
ModelIFR 100
CategoryThermometer
LanguageEnglish

Related product manuals