21
Table 3
Guidance and manufacturer’s declaration - electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment
Immunity Test IEC 60601
Test level
Compliance
level
Electromagnetic
environment - guidance
Conduced RF
IEC61000-4-6
150KHz to 80MHz:
3Vrms
6Vrms (in ISM
and amateur radio
bands)
80% Am at 1kHz
150KHz to
80MHz:
3Vrms
6Vrms (in ISM
and amateur
radio bands)
80% Am at
1kHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the 240391, including cables, than the
recommended separation distance calculated
from the equation appropriate for the frequency
of the transmitter. Recommended separation
distances:
d=1.2P;
d=2P