25
Table 5
Guidance and manufacturer’s declaration - electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or
the user of the device should assure that it is used in such an environment
Radiated RF
IEC61000-
4-3
(Test speci-
fications for
ENCLOSURE
PORT IM-
MUNITY to
RF wireless
commu-
nications
equipment)
Test
Fre-
quency
(MHz)
Band a)
(MHz)
Service a) Modulation b) Modula-
tion b)
(W)
Dis-
tance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385 380 –390 TETRA 400 Pulse
modulation b)
18 Hz
1,8 0,3 27
450 380 –390 GMRS 460,
FRS 460
FM c)
± 5 kHz devi-
ation
1 kHz sine
2 0,3 28
710 704 –
787
LTE Band 13,
17
Pulse
modulation b)
217 Hz
0,2 0,3 9
745
780