WARNINGS:
Single Use ONLY:
The STA Wand /SafetyWand/ handpiece and tubing assembly, as with any
syringe, opens a fluid pathway directly to the patient. This device is for single
use only. It must not be re-sterilized and must not be used on successive
patients or the same patient at a later visit. Reusing The STA Wand
/SafetyWand handpiece places the patient at risk. The anesthetic cartridge
must not be reused on multiple patients.
Do not bend needle during use:
Deformed or bent needles may interfere with the proper operation of the STA
Wand /SafetyWand
Lubricate the “O” Ring and Plunger:
A properly maintained and lubricated ‘O’ Ring is necessary for effective
functioning of the system. The following procedure is recommended:
a. Check ‘O’ Ring for cracks, deterioration, or lack of lubrication daily
b. If cracked or deteriorated, replace at once.
c. If dry, lubricate with silicon gel. While plunger is extended, lightly
lubricate plunger shaft with silicon gel. This will enhance smooth
performance.
STERILE, unless individual plastic package is opened or damaged.
CAUTIONS:
- Do not use the SafetyWand with needles longer than 1.25 inches
- Federal law restricts this device to sale by or on the order of a physician or
dentist.
- Maintain sterile conditions.