100
7−5−3. Cancellation for compensation by zero tracking
Cancellation for compensated data by zero tracking can be executed with the function F−98.
● When changing the target for zero tracking is required, be sure to
cancel the compensated portion by zero tracking once using the
function F−98.
● When the load display is a gross weight, the zero tracking of this unit
becomes effective. Therefore, the zero tracking does not work at the
time of the A/Z ON.
7−6. Stabilization filter
This instrument is equipped with the stabilization filter as a function for a digital filter to work
strongly when the state that change width of load indicated value is within the set range continues
longer than the fixed time.
7−6−1. What is the Stabiliz ation filter?
The stabilization filter is the function that a digital filter set by F−15 becomes effective when
the state continues longer than the time set by F−16 that the change width of load indicated
value is within the range set by F−17.