120
8−2. Function of Function data
D F−01 Selection of decimal point at display position
Initial value=00000 00000=Non
00001=10
1
00002=10
2
00003=10
3
00004=10
4
D F−02 Selection of A/D sampling rate
Initial value=00004 00000:10 times/s
00001:20 times/s
00002:50 times/s
00003:100 times/s
00004:200 times/s
D F−03 Selection of display rate
Initial value=00001 00000=4 times/s
00001=20 times/s
00002=50 times/s
00003=100 times/s
D F−04 Setting digital filter
Initial value=00001 Setting range:=00000〜00006
※Select the average times for digital filter. When the figure grows larger, the filter
becomes stronger, then effects from vibrations and so on are shown scarcely on the
display. However, if too large figure is selected, the response to variation of input
sensor will become worse.
※Using the digital filter where vibrations and so on may exist is effective for removing
the deflection on the display.
D F−05 Setting analog filter
Initial value=00001 00000=2Hz
00001=4Hz
00002=6Hz
00003=8Hz
00004=10 Hz