127
D F−90 Increment value (For reference)
※The increment value set when the calibration is applied can be referred to.
※The setting can’t be changed in the function.
D F−91 The maximum display value (For reference)
※The maximum display value set when the calibration is applied can be referred to.
※The setting can’t be changed in the function.
D F−92 The actual load value (For reference)
※The actual load value set when the calibration is applied(LCAL) can be referred to.
※When the calibration except LCAL is made, this value will not change.
※The setting can’t be changed in the function.
D F−93 Zero calibration value (For reference)
※The input voltage value that has read as the initial load value at the time of executing
calibration can be referred to.
※The setting can’t be changed in the function.
D F−94 Span calibration value (For reference)
※The input voltage value at the time of the maximum display can be referred to.
※The setting can’t be changed in the function.
D F−97 Prohibition of calibration
Initial value=00000 00000=Possible to calibrate
00001=Prohibition from calibration
D F−98 ZERO clear
Zero compensated data by zero set function can be cancelled.
When the
key is pressed with “F−98” displayed, “ZCLR” can be displayed.
(At the same time, the display lights on and off.)
At this point, press the
key when suspending the ZERO clear is desired.
Measurement mode can be returned and Zero clear will not be executed.
When the
key is pressed while “ZCLR” display lights on and off, “F−98” display can
be returned. Now, ZERO clear has completed.