126
D F−59 Setting the delay time of returning RS−485
Initial value=00005 Setting range:00000〜00999
Unit 1 ms
The delay is 999 ms at the “00999” setting.
D F−65 Setting the detection of stability range
Initial value=00010 00000:Detection of stability OFF
Setting range:00000〜00099
Unit 0.5D
※Effective only when the value from 00001 to 00099 with the F−66
D F−66 Setting the detection of stability time
Initial value=00005 00000:Detection of stability OFF
Setting range:00000〜00099
Unit 0.1s
※Effective only when the value from 00001 to 00099 with the F−65
D F−72 Setting the effective time for external control input
Initial value=00000 00000:50 ms
00001:20 ms
00002:10 ms
00003:5ms
00004:2ms
D F−84 Setting the occupied station number of CC−LINK
(Effective when the option is installed)
Initial value=00002 00000:1 station
00001:2 stations
00002:4 stations
● Change the setting of the occupied station number is corresponded to
the software of the instrument after the ROM Ver1.800.
The occupied station number is fixed as 4 before the ROM Ver1.700.
D F−85 Setting the station number of CC−LINK
(Effective when the option is installed)
Initial value=00001 Setting range:00001〜00061
D F−86 Setting the baud rate of CC−LINK
(Effective when the option is installed)
Initial value=00000 00000:156 kbps
00001:625 kbps
00002:2.5 Mbps
00003:5 Mbps
00004:10 Mbps
D F−87 Setting of the numeric expression of minus
Initial value=00000 00000:Expression of standard binary
(−1=FFFFFFFFH)
00001:At minus, The most significant digit is fixed to “8”.
(−1=80000001H)