135
Procedures
3
Press the key.
The load display shows “ALOW”.
4
Press the key.
The load display flashes on and off showing
“ALOW”.
At the same time, the analog output equivalent
to the minimum value of analog output is
obtained.
Adjust so the output will become the minimum
value with the right keys.
Minimum
value
Current output
(Optional)
Voltage output
(Optional)
0V4mA
By pressing the key continuously,
increment can be provided continuously
key :Decrease the analog output
key :Increase the analog output
5
Fine adjustment on the maximum value of
analog output
Press the key.
The display section shows as “A_HI”.