134
9−1−7. Fine adjustment 1 on analog output
Fine adjustment described here, is the one to arrange each “the minimum value” and “the
maximum value” without applying the actual load during the procedures. Refer to the
paragraph 9−1−8, for the fine adjustment with actual load applied.
● During the application on fine adjustment, if you want to suspend,
press the key. The minimum value data, the maximum value
data are kept as they were before entering the fine adjustment, and
the Measurement mode can be re−entered.
● Make fine adjustment one (1) hour or so after feeding the power.
You can make fine adjustment with safer conditions .
Procedures
1
Press the key for about one second.
The load display shows “FUNC”.
2
Press the key 9 times.
The load display changes as “FUNC”
→“CCAL”→“ACAL”→“LCAL”→“ZERO”
→“SPAN”→“TARE”→“CHEK”→“MONT”
→“VCAL”.