137
9−1−8. Fine adjustment 2 on analog output
The fine adjustment explained in this paragraph is the procedures with applying the actual
weight.
Warning ● Before making the fine adjustment, be sure to make scaling for the
analog output by referring to the paragraph 9−1−6.
If neglected, deviation of output can’t be adjusted during the fine
adjustment.
● During the application on fine adjustment, if you want to suspend the
fine adjustment, press the key. The zero data and the span data
are kept as they were before entering the fine adjustment, and the
Measurement mode can be re−entered.
● Make fine adjustment in one (1) hour or so after feeding the power.
You can make fine adjustment with safer conditions .
Procedures
1
Press the key for one second.
The load display shows “FUNC”.
2
Press the key 10 times.
The load display will change as “FUNC”
→“CCAL”→“ACAL”→“LCAL”→“ZERO”
→“SPAN”→“TARE”→“CHEK”→“MONT”
→“VCAL”→“VADJ”.