153
Function
No.
Operation Set value
09 Setting of time width for zero tracking (Contents of F−09)
00000=Zero tracking OFF
Setting range:00000〜00099
Unit:0.1 s
00000〜00099
15 Setting of the digital filter for the stabilized filter
(Contents of F−15)
00000=Stabilized filter OFF
Setting range:00000〜00006
※Strength of the digital filter for the stabilization filter is
selected. If the figure grows, the influence such as the
vibrations will not appears easily on the display because
the filter comes strongly.
※Effective only when 00001 to 00999 is set with F−16
and 00001 to 00999 is set with F−17.
00000〜00006
16 Setting of the time width for the stabilized filter
(Contents of F−16)
00000=Stabilized filter OFF
Setting range:00000〜00999
Unit:0.01 s
Time width:9.99s at the setting of “00999”
※Effective only when 00001 to 00006 is set with F−15
and 00001 to 00999 is set with F−17.
00000〜00999
17 Setting of the data width for the stabilized filter
(Contents of F−17)
00000=Stabilized filter OFF
Setting range:00000〜00999
Unit:1D
Data width:999D at the setting of “00999”
※Effective only when 00001 to 00006 is set with F−15
and 00001 to 00999 is set with F−16.
00000〜00999
20 Setting of the target for the analog output
(Contents of F−20)
00000:Gross weight
00001:Net weight
00000〜00001
21 Display value at the minimum value for analog output
(Contents of F−21)
Setting range:−99999 to 99999 Unit:1 count
−99999〜99999
22 Display value at the maximum value for analog output
(Contents of F−22)
Setting range:−99999 to 99999 Unit:1 count
−99999〜99999
35 Setting of the data width for the comparator hysteresis
(Contents of F−35)
00000=Hysteresis data width OFF
Setting range:00000〜00099
Unit:1D
Data width:99D at the setting of “00099”
00000〜00099
36 Setting the time width for comparator hysteresis
(Contents of F−36)
00000=Hysteresis time width OFF
Setting range:00000〜00999
Unit:0.01 s
Data width:9.99 s at the setting of “00999”
00000〜00999