152
5
Change of condition(Host→CSD−891B)
Command No.
Operation
50 Zero set
51 A/Z
52 A/Z OFF
Return(CSD−891B→Host)
246
@ 0 0 5 0 Terminator
Command No.
6
Writing comparative data(Host→CSD−891B)
2 4 6 8 10 12 14
@ 0 0 6 0 Terminator
Command No.
Sign
Set value
Command No. Operation
60 Writing S0 data
61 Writing S1 data
62 Writing S2 data
63 Writing S3 data
64 Writing S4 data
Return(CSD−891B→Host)
2468101214
@ 0 0 6 0 Terminator
Command No.
Sign
Set value
D The set value enters from the right end.
D Setting range is from −99 000 to 99 999.
D Never add the decimal point.
7
Writing the function data(Host→CSD−891B)
2 4 6 8 10 12 14
@ 0 0 7 0 0 4 Terminator
Function No.
Command No.
Sign
Set value
Function
No.
Operation Setting value
04 Setting of digital filter (Contents of F−04)
Setting range:=00000〜00006
00000〜00006
05 Setting of analog filter (Contents of F−05)
Setting range:=00000〜00004
00000〜00004
08 Setting of data width for zero tracking (Contents of F−08)
00000=Zero tracking OFF
Setting time:00000〜00099
Unit:0.5D
Data width of 49.5 D at the setting of “00099”
※Effective only when 00001〜00099 by F−09 is set.
00000〜00099
246
@ 0 0 5 0 Terminator
Command No.