246
@ 0 0 4 0 Terminator
Command No.
246
@ 0 0 4 1 Terminator
Command No.
151
Return(CSD−891B→Host)
2 4 6 8 10 12 14
@ 0 0 3 0 Terminator
Command No.
Sign
Set value
D The load data enters from the right end.
D When the data is minus, “−” sign and when plus, “+” sign is added.
D The sign data performs zero suppress.
D In case that the decimal point is set with the F−56, and at the same time the
decimal point is specified with the F−01, it will be added to the specified
position.
D The empty sections are all spaces.
3
Reading−out the condition (Host→CSD−891B)
Command No.
Operation
40 Reading out the condition
Return(CSD−891B→Host)
2 4 6 8 10 12 14
@ 0 0 4 0 a b c d e Terminator
Condition
Command No. Space
a:RUN LED display “1”=ON, “0”=OFF
b:A/Z LED display “1”=ON, “0”=OFF
c:LOCK LED display “1”=ON, “0”=OFF
d:HOLD LED display “1”=ON, “0”=OFF
e:CHECK LED display “1”=ON, “0”=OFF
4
Reading out the comparative results (Host→CSD−891B)
Command No.
Operation
41 Reading out comparative result
Return(CSD−891B→Host)
2 4 6 8 10 12 14
@ 0 0 4 1 S0 S1 S2 S3 S4 Terminator
Condition
Command No. Space
S0: “1”=ON, “0”=OFF
S1: “1”=ON, “0”=OFF
S2: “1”=ON, “0”=OFF
S3: “1”=ON, “0”=OFF
S4: “1”=ON, “0”=OFF