246
@ 0 0 0 1 Terminator
Command No.
156
Function No.
Operation
44 Reading out the setting value of the BCD P.C. output logic (Contents of F−44)
45 Reading out the setting value of BCD P.C. width (Contents of F−45)
46 Reading out the setting value of BCD output times (Contents of F−46)
50 Reading out the setting value of RS−232C operation mode (Contents of F−50)
51 Reading out the setting value of RS−232C stream mode (Contents of F−51)
52 Reading out the setting value of RS−232C/422/485 baud rate (Contents of F−52)
53 Reading out the setting value of RS−232C/422/485 data bit length
(Contents of F−53)
54 Reading out the setting value of RS−232C/422/485 stop bit (Contents of F−54)
55 Reading out the setting value of RS−232C/422/485 terminator (Contents of F−55)
56 Reading out the setting value of the decimal point of RS−232C/422/485
transmission code (Contents of F−56)
57 Reading out the setting value of ID number for RS−422/485 (Contents of F−57)
58 Reading out the setting value of RS−422/485 changeover (Contents of F−58)
59 Reading out the setting value of the delay times for RS−485 transmission
(Contents of F−59)
65 Reading out the detection of stability range (Contents of F−65)
66 Reading out the detection of stability time (Contents of F−66)
72 Reading out the setting value of the effective time for external control input
(Contents of F−37)
90 Reading out the increment value (for reference) (Contents of F−90)
91 Reading out the maximum display value (for reference) (Contents of F−91)
92 Reading out the actual load value (for reference) (Contents of F−92)
93 Reading out the zero calibration value (for reference) (Contents of F−93)
94 Reading out the span calibration value (for reference) (Contents of F−94)
97 Reading out the setting value of prohibiting calibration (Contents of F−97)
Return(CSD−891B→Host)
2 4 6 8 10 12 14
@ 0 0 7 1 0 1 Terminator
Function No.
Command No.
Sign
Set value
(3) Calibration Command mode
1
Start of calibration mode(Host→CSD−891B)
Command
No.
Operation
01 Calibration to register the output of
strain gage applied transducer at the
maximum display after setting the
load to zero
02 Calibration to register the output of
strain gage applied transducer at
zero and the maximum display
03 Calibration to register read the
output of strain gage applied
transducer read after setting the load
to zero or actual load.
04 Zero fine adjustment