32
Procedures
Press the key.
The load display shows ”0.0000”, and the digit
of 10
0
will flash on and off. In case that
calibration has completed already, the registered
output value of load cell at that time is
displayed. Set the output value for load cell with
the initial load application with the right keys.
Though the number of digits has not
prepared in the ”Inspection data” for load
cell so many as shown in the right figure,
extra digits are necessary for the
compensat
on w
t
t
stan
ar
po
nt at
internal of the instrument.
In case of actual setting, insert “0”, into the
extra digits.
As for the value for extra digit, when tare
compensation and fine adjustment on load
are applied, it is rewritten as a compensated
value automatically.
Setting range for the output of load cell is
from −2.500 0 mV/V to 2.500 0 mV/V.
By pressing the key continuously,
increment can be provided continuously.
key :Set value carry key
key :Set value inclement key
key :Set value initialization key
9
Press the key.
The load display shows “S MV”.