39
Procedures
Press the key.
The “ZERO” on load display flashes on and off,
and zero adjustment can be started.
10
Warning : Take care not to apply load
variations due to vibrations and so on.
If load variation is applied, the zero point
will not stabilized, in due course there is a
possibility that correct reading of zero
won’t be obtained.
When completed, the display on the load display
shows “SPAN”.
However, when the initial load is not entered the
range of −2.5 mV/V to 2.5 mV/V, the right Error
code is shown for about 2 seconds, then the
display on the load display section is shown as “
ZERO”, and then the step 9 can be entered.
Error code
TE−L:Zero point −OVER
TE−H:Zero point +OVER
Error code
11
Apply th
e
same load on th
e
load cell as se
t
in th
e
step 8.