5−3. Calibration method by communication
Warning ● Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration.
If calibration shall not be made, correct measured results may not be
obtained nor may cause malfunction in the instrument and there may
exist damage in peripheral equipments.
Besides, even though calibration has been made, there may occur the
similar case when the result is not correct, so make calibration again.
● During the calibration is executing, be sure to set Tare weight
cancellation clear, and to make cancellation (Execution of F−98) for
compensated data on zero set and set the OFF position of Zero
tracking(Setting “00000” on F−08 and F−09). If neglected, it returns
the error command (Error command No.02) to the host.
● Also In case that the command not suitable for the procedure is
transmitted during the calibration, the error command(Error
command No.02) is send back to the host side.
● The calibration of this instrument is not interlocked with the display
value at the minimum analog output (F−21) or at the maximum
analog output (F−22). Make sure to execute or confirm the setting by
F−21 and F−22 when the calibration has been made. If neglected,
the correct measured results may not be obtained nor may cause
malfunction in the instrument and there may exist damage in
peripheral equipments.
● The calibration by the communication is possible when the optional
RS−232C interface or RS−422/485 interface is mounted.