5−3−1. Calibration method by communication to register the output of strain gage applied
transducer at the time of maximum display after setting the load to zero.
Warning ● Before using a new instrument or exchanging the strain gage applied
transducer for a new one, be sure to make calibration.
If calibration shall not be made, correct measured results may not be
obtained nor it may cause malfunction in the instrument and there
may exist damage to the peripheral equipment.
Besides, even though the calibration has been made, there may occur
the similar case when the result is not correct, so make calibration
again.
● During the calibration is executing, be sure to make available for the
calibration set (Setting “00000” on F−97). Tare weight cancellation
clear, and to make cancellation (Setting of F−98) for compensated
data on zero set and set the OFF position of Zero tracking(Setting “
00000” on F−08 and F−09). If neglected, it returns the error
command (Error command No.01) to the host.
● The error command (Error command No.02) similarly returns to the
host when the command not suitable for the procedure is transmitted
during the calibration.
● Please match function No.F−50 to F−59 as to the communication to
host’s specification and change (Refer to the paragraph 8, 9−3 and
9−4). If neglected, the communication may not be executed correctly.