61
Procedures
Transmit the change command data of span
mV/V value from the host, and set the output
value corresponding to the maximum display
value set in step 3.
The value set here set the value whose 0.200 0
mV/V ( “02000” on the actual setting value) is
bigger than the value set in step 4.
The display in the load display section displays
“−RS−” and “END” alternately.
Warning : When the inspection data
sheet of the load cell indicates only the
digit of “X.XXXX”, put “0” for the extra
digit.
The numeric value of an extra digit, etc.,
is automaticall
rewritten as th
2 4 6 8 10 12 14
@ 0 0 1 5 0 3 0 0 0 Termi
Return CSD−891B→Host
・Normal termination
2 4 6 8 10 12 14
@ 0 0 1 5 0 3 0 0 0 Termi
Command Host→CSD−891B
Output value of
load cell
ID No.
ID No.
-nator
Output value of
-nator
compensated value after executing the
tare weight compensation or fine
adjustment of the load.
Don’t attach the decimal point to the
setting value in the command data.
The setting value in the command data is
put from the right, and an unnecessary
space is given as space.
2 4 6 8 10 12 14
@ 0 0 1 5 S P - L Termi
・Less the 0.200 0 mV/V
Error code
ID No.
2 4 6 8 10 12 14
@ 0 0 1 5 S P - H Termi
・Higher than 3.100 0 mV/V
Error code
ID No.
load cell
-nator
-nator
Transmit the termination command from the
calibration mode from the host.
The calibration data is made effective and the
calibration mode is over.
The display in the load display section displays
the present load value.
246
@ 0 0 1 9 Terminator
Command Host→CSD−891B
ID No.
2 4 6 8 10 12 14
@ 0 0 1 9 E N D Termi
Return CSD−891B→Host
ID No.
-nator