60
Procedures
Transmit the change command of zero mV/V
value from the host, and set the output value
from load cell in the initial load condition. The
display in the load display section displays
“−RS−” and “S MV” alternately.
2 4 6 8 10 12 14
@ 0 0 1 4 + 0 0 0 0 0 Termi
ID
No
Command Host→CSD−891B
Sign(
+
/
)
-nator
4
When the inspection data sheet of load
cell indicates only the digit of “X.XXXX”,
put “0” for the extra digit. When the tare
weight compensation and the fine
adjustment of the load is done, the
numerical value of an extra digit, etc., is
automatically rewritten in correction.
Though the setting range of load cell is
from −2.500 0 mV/V to 2.500 0 mV/V,
the range of the actual setting value is “
−25000 to 25000” because the decimal
point is not added.
Don’t attach the decimal point to the
setting value in the command data.
The setting value in the command data is
put from the right, and an unnecessary
space is given as space.
ID
N
o.
2 4 6 8 10 12 14
@ 0 0 1 4 + 0 0 0 0 0 Termi
ID No.
Return CSD−891B→Host
・Normal termination
2 4 6 8 10 12 14
@ 0 0 1 4 T E - L Termi
ID No.
・Less than −2.500 0 mV/V
Output value of load cell
2 4 6 8 10 12 14
@ 0 0 1 4 T E - H Termi
・Higher than 2.500 0 mV/V
Sign(
+
/
−
)
Sign(+/−)
Output value of load cell
-nator
-nator
-nator