65
Procedures
Transmit the change command data of the actual
load from the host, and set the load value on load
cell. Set the load value on load cell less than the
maximum display value set in step 3, and the
maximum value which can load in the range of
(The minimum digits×100) 〜 99 999.
The display in the load display section displays
“−RS−” and “ZERO” alternately.
2 4 6 8 10 12 14
@ 0 0 1 2 2 0 0 0 Termi
ID No.
Command Host→CSD−891B
2 4 6 8 10 12 14
Return CSD−891B→Host
・Normal termination
Actual load value
-nator
4
Don’t attach the decimal point to the
setting value in the command data.
The setting value in the command data is
put from the right, and an unnecessary
space is given as space.
@ 0 0 1 2 2 0 0 0 Termi
ID No.
2 4 6 8 10 12 14
@ 0 0 1 2 E R - 2 Termi
ID No.
・Set in error
Actual load value
-nator
-nator
Set in the initial load condition.
Transmit the start command of the calibration
at zero point of “
Calibration method to
register by reading output value of strain
gage applied transducer in the conditions of
zero/actual load application individually.
”.
The instrument transmit the return signal, and
the zero adjustment is started.
The display in the load display section displays
246
@ 0 0 1 3 Terminator
Command Host→CSD−891B
Return CSD−891B→Host
ID No.
“−RS−” and “ZERO” alternately.
Warning : Take care not to apply load
variations due to vibrations and so on.
If load variation is applied, the zero point
will not stabilized, in due course there is a
possibility that correct reading of zero
won’t be obtained.
2 4 6 8 10 12 14
@ 0 0 1 3 Z E R O Termi
-nator