EasyManua.ls Logo

Minebea CSD-891B - Page 81

Default Icon
230 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
67
Procedures
8
Transmit the command for “Status reading out
in the calibration” from the host.
According to the return data, proceed to the
following steps.
・When the return data is “END”
As termination is in normal, you can proceed to
the next step.
・When the return data is “SP−L”
The load is less than 0.2 mV/V.
After setting the load within the range from 0.2
mV/V to 3.1 mV/V, proceed to the step 7.
・When the return data is “SP−H”
The load is exceeding over 3.1 mV/V.
After setting the load within the range from 0.2
mV/V to 3.1 mV/V, proceed to the step 7.
・When the return data is “ZERO”
Now under the calibration.
Until the return data becomes except “ZERO”,
repeat to transmit the command of “Status
reading out in the calibration” from the host.
The display in the load display section displays
“−RS−” and “SPAN” alternately.
2 4 6 8 10 12 14
@ 0 0 0 9 E N D Termi
ID No.
Command CSD−891B→Host
・Normal termination
2 4 6 8 10 12 14
@ 0 0 0 9 S P - L Termi
ID No.
2 4 6 8 10 12 14
@ 0 0 0 9 S P - H Termi
ID No.
2 4 6 8 10 12 14
@ 0 0 0 9 S P A N Termi
ID No.
246
@ 0 0 0 9 Terminator
Command Host→CSD−891B
ID No.
・Low input
・High input
・Under the calibration
-nator
-nator
-nator
-nator
9
Transmit th
e
termination command of th
e
calibration mode from the host.
The calibration data is made effective and the
calibration mode is over.
The display in the load display section displays
the present load value.
246
@ 0 0 1 9 Terminator
Command Host→CSD−891B
2 4 6 8 10 12 14
@ 0 0 1 9 E N D Termi
Return CSD−891B→Host
・Normal termination
ID No.
ID No.
-nator