5−3−4. Zero fine adjustment by communication
Warning ● During the calibration is executing, be sure to make available for the
calibration set (Setting “00000” on F−97). Tare weight cancellation
clear, and to make cancellation (Setting of F−98) for compensated
data on zero set and set the OFF position of Zero tracking(Setting “
00000” on F−08 and F−09). If neglected, it returns the error
command (Error command No.01) to the host.
● The error command (Error command No.02) similarly returns to the
host when the command not suitable for the procedure is transmitted
during the calibration.
● Please match function No.F−50 to F−59 as to the communication to
host’s specification and change (Refer to the paragraph 8, 9−3 and
9−4). If neglected, the communication may not be executed correctly.
● This calibration method is possible when the optional RS−232C
interface or the RS−422/485 interface is mounted.
● When RS−232C interface is used, the ID No. becomes “00”.
● During the calibration, to interrupt the calibration “Calibration mode
interruption command” is transmitted, or press the key.
The calibration data is kept as before entering the calibration and
returns to the measurement mode.