75
Procedures
1
Transmit the calibration mode starting
command from the host.
The display in the load display section displays
“−RS−” and “ZERO”
246
@ 0 0 0 6 Terminator
Command Host→CSD−891B
2 4 6 8 10 12 14
@ 0 0 0 6 T A R E Termi
ID No.
Return CSD−891B→Host
ID No.
-nator
Here, set the initial load condition.
Transmit the start command of “Calibration
procedure to apply registration again for zero
point only” from the host. Transmit the return
data from this instrument, and tare weight
cancellation is started.
The display in the load display section displays
“−RS−” and “ZERO”
246
@ 0 0 1 3 Terminator
Command Host→CSD−891B
ID No.
As for the load value read by the load data
reading out command, the decimal point
adheres to the decimal point position in
F01 when you set to add the decimal point
in F−56.
2 4 6 8 10 12 14
@ 0 0 1 3 Z E R O Termi
Return CSD−891B
Host
-nator