76
Procedures
3
Transmit the command of “Reading out in the
calibration” from the host.
According to the return data, proceed to the
following step.
・When the return data is “END“
As the termination is in normal, proceed to the
next step.
・When the return data is “TE−L”
The initial load is less than −2.5 mV/V.
After setting the initial load from −2.5 mV/V to
2.5 mV/V, proceed to the step 2.
・When the return data is “TE−H”
The initial load exceeds over 2.5 mV/V.
After setting the initial load from −2.5 mV/V to
2.5 mV/V, proceed to the next step.
・When the return data is “ZERO”
Until the return data becomes except “ZERO”,
repeat to transmit the command of “Status
reading out in the calibration” from the host.
The display in the load display section displays
“−RS−” and “SPAN” alternately.
2 4 6 8 10 12 14
@ 0 0 0 9 E N D Termi
ID No.
Command CSD−891B→Host
・Normal termination
2 4 6 8 10 12 14
@ 0 0 0 9 T E - L Termi
ID No.
2 4 6 8 10 12 14
@ 0 0 0 9 T E - H Termi
ID No.
2 4 6 8 10 12 14
@ 0 0 0 9 Z E R O Termi
ID No.
246
@ 0 0 0 9 Terminator
Command Host→CSD−891B
ID No.
・When the input is low
・When the input is high
・Under the calibration
-nator
-nator
-nator
-nator
4
Transmit th
e
termination command of th
e
calibration mode from the host.
The calibration data is made effective and the
calibration mode is over.
The display in the load display section displays
the present load value.
246
@ 0 0 1 9 Terminator
Command Host→CSD−891B
2 4 6 8 10 12 14
@ 0 0 1 9 E N D Termi
Return CSD−891B→Host
・Normal termination
ID No.
ID No.
-nator