34
- the above phenomenon does not affect the basic safety and essential
performance of the device, and the user can use it according to the
instruction. If you want to avoid the above phenomenon, please use it
according to the environment specified in the manual.
Declaration - electromagnetic emission
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of device should assure that it is used in such
an environment.
Emissions test Compliance
Electromagnetic
environment - guidance
RF emissions
CISPR 11
Group 1
The device uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for
domestic establishment
and in establishment
directly connected to
the public low-voltage
power supply network that
supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of device should assure that it is used in such
an environment.
Immunity test
IEC 60601test
level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
±8 kV
contact
±2 kV, ±4 kV,
±8 kV, ±15
kV air
Floors should be wood,
concrete or ceramic tie.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.