35
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines ±
1 kV for input/
output lines
Not
applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 0.5kV, ± 1 kV
line(s) to lines
± 0.5kV, ± 1 kV,
± 2 kV line(s) to
earth
Not
applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-
4-11
0 % UT; 0.5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270°and 315°
0 % UT; 1 cycle
and
70 % UT; 25/30
cycles
Single phase:
at 0°
0 % UT;
250/300 cycles
Not
applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the device
requires continued
operation during power
mains interruptions, it
is recommended that
the device be powered
from an uninterruptible
power supply or a
battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m 30 A/m
Power frequency mag-
netic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
ENGLISH