38
Emissions test Compliance
Electromagnetic
environment - guidance
RF emissions
CISPR 11
Group 1
The device uses RF energy
only for its internal function.
Therefore, its RF emissions
are very low and are not likely
to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The device is suitable for
domestic establishment and
in establishment directly
connected to the public low-
voltage power supply network
that supplies buildings used
for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
uctuations/
icker emissions
IEC 61000-3-3
Not applicable
Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment
specied below. The customer or the user of device should assure that it
is used in such an environment.
Immunity
test
IEC
60601test
level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15
kV air
±8 kV contact
±2 kV, ±4 kV,
±8 kV, ±15
kV air
Floors should be
wood, concrete or
ceramic tie. If oors are
covered with synthetic
material, the relative
humidity should be at
least 30%.
ENGLISH