Serviceanleitung
Operationsmikroskop
MÖLLER Hi-R 1000
Service Manual
Operating Microscope
MÖLLER Hi-R 1000

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Moller Hi-R 1000 and is the answer not in the manual?

Summary

Important Notes

WARRANTY AND LIMITATION OF LIABILITY

Troubleshooting

Error message NO MICROSCOPE

Indicates microscope not connected or power supply issue.

Error message HANDGRIP SENSOR

Relates to sensor function indication via LED V4 on the main PCB.

Error message MICRO.SERVO-DRIVE

Indicates drives are switched off due to main processor or regulator PCB issue.

Error message MICRO. HANDSWITCH

Concerns errors related to stepping motor controls and hand switch functions.

Error message MOTOR-STOP-MODE

Activated when motor stop switch is pressed, interrupting motion controller power.

One of the microscope functions does not work

Points to defective stepping motor control, motor, or motor cable.

Microscope cannot be operated via stand or foot switch

Indicates a defective microscope interface if hand switch operation is normal.

No microscope function

Covers missing power supply or watchdog function issues with the main processor.

Servo drive is self-acting

Suggests issues with regulator PCB, motion controller, or servo motor.

Exchange of Assemblies

Main PCB

Instructions for removing and installing the main printed circuit board.

Function Test