EasyManua.ls Logo

Moore Industries TRZ - Page 7

Default Icon
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The Interface Solution Experts 3
TRZ
Input Type
α
Measuring
Range
Minimum
Span
Input
Accuracy
RTD
Platinum
0.003850
25-500
200 to 850°C
328 to 1562°F
10°C
18°F
±0.2°C
±0.36°F
1000
200 to 320°C
328 to 608°F
0.003902
25-500
100 to 650°C
148 to 1202°F
1000
100 to 320°C
148 to 608°F
0.003916
25-500
200 to 500°C
328 to 932°F
1000
200 to 320°C
328 to 608°F
Nickel 0.00618
25-500
50 to 250°C
58 to 482°F
10°C
18°F
±0.2°C
±0.36°F
1000
Copper 0.00427 25-1000
50 to 200°C
58 to 392°F
10°C
18°F
±0.2°C
±0.36°F
Direct
Resistance or
Potentiometer
n/a n/a
0 to 390
5
±0.08
0 to 2200
25
±0.4
Input Type
Maximum
Range
Conformance
Range
Minimum
Span
Input
Accuracy
T/C
J
210 to 1200°C
346 to 2192°F
200 to 1000°C
328 to 1832°F
50°C
90°F
±1.0°C
±1.8°F
K
250 to 1372°C
454 to 2502°F
230 to 1370°C
382 to 2498°F
E
270 to 1000°C
454 to 1832°F
250°C to 900°C
418 to 1652°F
T
250 to 400°C
454 to 752°F
220 to 400°C
364 to 752°F
40°C
72°F
R
50 to 1768°C
58 to 3216°F
0 to 1750°C
32 to 3182°F
100°C
180°F
±2.0°C
±3.6°F
S
N
270 to 1300°C
454 to 2372°F
130 to 1300°C
202 to 2372°F
50°C
90°F
±1.0°C
±1.8°F
B
100 to 1820°C
392 to 3308°F
500 to 1820°C
932 to 3308°F
50°C
90°F
±2.0°C
±3.6°F
C (W5)
0 to 2300°C
32 to 4172°F
0 to 2300°C
32 to 4172°F
100°C
180°F
D (W3)
L
200 to 900°C
328 to 1652°F
200 to 900°C
328 to 1652°F
50°C
90°F
±1.0°C
±1.8°F
U
200 to 600°C
328 to 1112°F
200 to 600°C
328 to 1112°F
mV DC
10 to +70mV
n/a
2mV ±0.04mV
0.1 to +1.1V 20mV ±0.4mV
Table 1. Input Types and Accuracy Specifications for the TRZ Transmitter